GE H
EALTHCARE
D
IRECTION
DOC0500644, R
EV
4 D
ISCOVERY
, L
IGHT
S
PEED AND
B
RIGHT
S
PEED
DICOM C
ONFORMANCE
S
TATEMENT
122 of 157
8.6 SUPPORT OF EXTENDED CHARACTER SETS
GSPS feature supports ISO_IR 100 character set.
8.7 ERROR HANDLING AND RECOVERY
Exam Application has an extended error codes and the standard error codes are also
handled. If the sending GSPS/PPS objects request has failed or response is not received
before the association timeout, the operation is deemed ‘Failed’ and the system updates
the state and indicates the same on the user interface.
If the operation is ‘Failed’, detailed message is logged into system log-file and system
provides an alternative mechanism to retry the failed operation through the user interface
to ensure that transient failures do not affect the feature performance.